es8 datalogger
Low cost data recorder for vibration measurement and an 8 channel AD converter with a USB interface for laboratory use.
es8 Datalogger - Specification
| Number of channels | 8 channels. Individually ON/OFF selectable. |
| A/D Converter | 16 bits successive approximation type. All channels simultaneously sampled (input filter not included) |
| Input Range/Type | ±2, 5 and 10V / unbalance Impedance: 1M Ohm |
| Sampling Frequency |
5KHz-1/60Hz (same sample frequency applied to all channels) |
| Record Mode | Manual, Interval, Level Trigger (Edge/Window), External Trigger, Pre/Post Trigger |
| Comparator Output |
One relay contact (normally open), Relay rating 40V 0.25A |
| Display | 20 characters x 4 lines Character type LCD with backlight |
| Operation Key | REC, STANDBY, STOP, MENU, ENTER, ESC, UP, DOWN, RIGHT, LEFT |
| Record Media | Compact Flash (FAT16) Maximum 2GB |
| Data Format |
Text CSV format or binary TAFFmat format selectable. The maximum sampling frequency is 50Hz for the CSV format. |
| PC Connection | USB 2.0 Full speed only. * USB Bus power operation is not available. |
| Clock Calibration | 30 second adjustment |
| Operating Temperature and Humidity |
-10°to +50° (In use under low temperature, LCD may dew) 10% to 85% (non condensing) |
| Power Supply |
Internal : AA type Nickel Metal-Hydride or 4 Alkaline batteries External: 8-15V DC 12V Please use the capacity of more than 0.2A for the DC operation, or the accessory AC adaptor. |
| External Dimension | Approx. 150W x 100D × 40H mm |
| Weight | Approx. 400g (internal batteries not included) |
| Accessories |
User Manual, AC adaptor, D-sub 25pin connector, Navi software (for real-time display and control), AA Type dry cell batteries. |
| Safety Standard | VCCI Class A |
| Option |
Input cable (D-sub 25 pin -BNC x 8) Terminal box |
|
* Exact product specifications and exterior appearance are subject to change without notice. (January, 2005) * The company name and the product name are registered trademarks. |
|